Refractive Universal IntegratorIOL power calculator

Disclaimer

The Refractive Universal Integrator (RUI) assists eye-care professionals in choosing intraocular lens (IOL) power for cataract surgery. Use it together with a full ophthalmic examination and appropriate diagnostic tests, and compare the result with your usual formula.

Its results are not medical or surgical instruction. The surgeon makes their own independent decision on IOL power and remains solely responsible for the treatment and refractive outcome of their patient.

Check every value against the original biometry, including any value read from a photographed printout.

RUI is provided as is, without warranty of any kind, express or implied, and is not registered as a medical device with the Health Sciences Authority (HSA) or any other regulator. You assume all risk of using it. The developer and anyone connected with this site accept no liability for any direct or indirect loss or damage arising from its use.

The calculator does not save patient records or calculation history. Optional Google sign-in uses a temporary account session, separate from patient and surgeon details entered for a calculation. See the privacy and sign-in policy.

RUI combines an advanced optical model with neural network training. It was designed for modern biometry, including a wide range of special scenarios, and draws on multiple biometric measurements known to influence IOL power.

RUI lets you select and revise the planned IOL so you can clearly communicate lens choices and changes to the wider surgical team responsible for planning, lens logistics and implantation.

The calculator and this site are the intellectual property of Dr SR Ng and may not be copied, reproduced, decompiled or reverse engineered. This disclaimer is governed by the laws of Singapore.

This disclaimer appears on every visit. Signing in does not replace acceptance of the disclaimer.

Privacy and sign-in policy

Acknowledgements

These surgeons have helped with contribution of data and/or refinement of usability and calculations. This work would not be possible without them.

We also thank the Singapore National Eye Centre Health Research Endowment Fund for its support.

Patient and eye

Eye
Age or birth year optional, recommended for greater accuracy (enter one)
years orborn

Eye status

Previous refractive surgery

For incisions of 2.75 mm or less, consider using an SIA of 0.1 D or lower.

Why can centroid SIA be lower?

Mean SIA averages the magnitudes of the changes in individual eyes. Centroid SIA averages both magnitude and direction. Changes in different directions can partly cancel, so the centroid magnitude can be lower than the mean SIA magnitude. With small clear corneal incisions, centroid SIA can be small, but it varies by surgeon and incision technique. A low centroid does not mean that every eye has little or no SIA.

For help calculating your centroid SIA, email the author with paired pre- and postoperative keratometry from at least 100 eyes, with patient names and identifiers removed.

Biometry

Off · Optical / immersion A-constant

D at°
D at°
Posterior or total K optional
D at°
D at°

Enter posterior corneal power as a positive magnitude.

Lens

Special cases

Keratoconus coming soon
Keratoconus
Previous refractive surgery

Refractive history is used for myopic LASIK / PRK only. Enter both values or neither. Measured posterior K or total K (under Posterior cornea) is used if entered, and corneas of K 33–50 D are accepted after refractive surgery.

Lens status and fixation phakic, in the bag unless changed here: pseudophakic, aphakic, scleral or iris

Scleral fixation needs white-to-white and does not use ACD. Iris fixation uses the lens's own A constant for its position, so enter the manufacturer's constant for the position you use. An aphakic eye needs scleral or iris fixation; measure its axial length with the biometer set to aphakic. Pseudophakic: enter the ACD measured to the front of the existing IOL (biometer set to pseudophakic). For an IOL exchange, the existing IOL power and current refraction refine the result if entered. For a piggyback lens, enter the current refraction.

Silicone oil temporary or permanent

Temporary (default): the oil is expected to be removed later. The power is for the eye after removal, and the result shows the refraction to expect while the oil is still in.
Permanent: tick this only if the oil will stay. The power is then compensated for the oil; if the oil is removed later, the eye becomes myopic. Choose the lens's optic shape and refractive index if you know them (hydrophobic acrylic about 1.47–1.55, PMMA 1.49). In both cases, measure axial length with the biometer's silicone-oil setting.

Fill from uploads

Your photo is read on this calculator’s engine and does not depend on third-party OCR vendors. It is sent over an encrypted connection through Cloudflare. The application processes it in memory without saving the photo or recognised content to files or logs.

RUI automatically selects suitable readings from your uploaded reports. You can review or change the source beside each measurement. Unreadable or unsupported fields stay empty.

IOLMaster, ARGOS, Tomey OA-2000, Lenstar LS 900, Pentacam and supported ultrasound reports have been tested.

Exactly how photo reading works
  1. Choose a photo. Upload image opens a file picker to choose photos from your device.
  2. Prepare. On your device the image is scaled down to an appropriate resolution.
  3. Send for reading. The image goes over an encrypted connection through Cloudflare to this calculator’s reading engine, hosted in Singapore. The application processes the image in memory and releases it after responding, without deliberately saving its content. Operational records contain request metadata and service diagnostics.
  4. Read the page. The OCR engine reads the photo several times: the whole page, the photo as taken, and enlarged views of the header, eye status and biometry areas.
  5. Check the readings. The engine checks extracted readings for consistency before filling supported fields. Uncertain readings stay empty. Use View source to compare the entries with the original report.
  6. Review source readings. Patient name, ID, birth year and sex, lens status, refractive surgery, AL, ACD, LT, WTW, CCT, K1/K2 and TK1/TK2 with their axes can be filled. Use View source to compare the image with the entries. Patient conflicts must be resolved or removed; measurement discrepancies appear as warnings without acknowledgement clicks.
  7. If the server reading service is unavailable, a simpler reader can run on your device. It supports selected IOLMaster 700 layouts and fills measurements only. Other reports may require manual entry.
  8. Calculate. Only measurement numbers, options, age and sex are sent for calculation, exactly as if you had typed them. Patient name, ID and birth year are not included in the calculation request.
    Not yet processed, please click Calculate